QAL2: what it is, when it is required and how to prepare the AMS calibration
QAL2 is the procedure used to calibrate an Automated Measuring System (AMS) for emissions on the actual plant. It is the second of the three quality assurance levels defined by EN 14181 and, together with the annual surveillance test (AST), it is the check on which the legal validity of the data reported to the competent authority depends.
In this article we look at what it involves, when it has to be repeated and how to prepare for it so that the test does not have to be done again.
What QAL2 is
An analyser certified to QAL1 measures correctly in the laboratory, but conditions at the stack are different: flue gas composition, moisture, temperature, sampling point location and sampling lines. QAL2 establishes the calibration function that links the AMS readings to the "true" values measured with the Standard Reference Method (SRM), and shows that the AMS uncertainty is within the limits set by legislation.
The test is carried out by a testing laboratory accredited to EN ISO/IEC 17025 for the reference methods used, independent of the operator. The operator, or whoever manages the AMS on its behalf, must make the system and its data available to the laboratory.
How the test is carried out
Parallel measurements
The laboratory performs a series of parallel measurements between the AMS and the reference method. EN 14181 requires at least 15 valid measurements, spread over at least three days and with plant operating conditions as representative of normal operation as possible. The wider the range of concentrations covered, the more reliable the calibration function.
During the measurements the AMS must run under normal conditions: no extraordinary maintenance, zero and span checks performed according to procedure, raw data acquired and stored.
Calibration function
The pairs of values collected are used to calculate the calibration function, usually a straight line of the form *y = a + b·x*. When the measured concentrations are all very close to each other, as often happens with stable plants and low emissions, the standard provides a specific procedure that also uses the analyser's zero readings.
The resulting function must be applied in the DAHS to the AMS values before averages are calculated and compared with the emission limit values.
Variability test
The variability test checks that the residual scatter between the calibrated AMS and the reference method is compatible with the maximum permissible uncertainty, expressed as a 95% confidence interval as a percentage of the emission limit value. For large combustion plants and waste incineration plants the percentages are set by Directive 2010/75/EU; for other plants the reference value is given by the permit, the AMS management manual or regional guidelines. If the test fails, the AMS cannot be considered compliant until the cause has been found and QAL2 repeated.
Valid calibration range
The calibration is only valid within the valid calibration range, which depends on the values measured during the test. If in normal operation the AMS often measures values outside this range, the standard requires the calibration to be extended or QAL2 to be repeated. That is why the DAHS should flag out-of-range values automatically.
When it has to be repeated
QAL2 is carried out:
- at commissioning of the AMS or after its installation
- at least every five years, or more often if the permit or sector legislation requires it
- after significant changes to the plant or fuel that alter the emission characteristics
- after replacement or a major repair of the analyser
- when the annual AST fails
- when measured values fall outside the valid calibration range too often
Actual deadlines must always be checked against your own permit: many Italian AIA and AUA permits set stricter frequencies or conditions than the standard.
QAL2, AST and QAL3: how they fit together
- QAL2 establishes the calibration function and verifies uncertainty.
- The AST (Annual Surveillance Test) checks every year, with a smaller number of parallel measurements, that the calibration function is still valid.
- QAL3 continuously monitors zero and span drift with control charts between one QAL2 and the next.
If QAL3 is not managed properly, analyser drift only shows up at the AST, with the risk of having to repeat QAL2 and justify the data for the previous period.
How to prepare and avoid repeating the test
Many failed QAL2 tests are not caused by the analyser but by the state the system is in when the test starts. Before the laboratory arrives it is worth:
- checking the AMS: heated lines, probe, filters, NO₂/NO converter, calibration gases within their validity date
- reviewing the QAL3 control charts for the last few months to spot drift or instability
- planning with plant operations representative and, where possible, varied load conditions
- making sure the DAHS records raw data at the required time resolution and is ready to receive the new calibration function
- updating the AMS management manual with the new calibration and the test results
How APS Automazione can help
APS Automazione maintains AMS before, during and after QAL2, including systems installed by other suppliers: preparing the instrumentation, supporting the laboratory during the parallel measurements and entering the calibration function into the DAHS. Our DAHS software handles QAL2 and QAL3 and flags values outside the valid calibration range.